Electromagnetic induction imaging: signal detection based on tuned-dressed optical magnetometry

Year: 2021

Authors: Bevilacqua G., Biancalana V., Dancheva Y., Fregosi A., Napoli G., Vigilante A.

Autors Affiliation: Univ Siena, Dipartimento Ingn Infornuzz & Sci Matemat DIISM, Via Roma 56, I-53100 Siena, Italy; Univ Siena, Dipartimento Sci Fis Terra & Ambience DSFTA, Via Roma 56, I-53100 Siena, Italy; Aerospazio Tecnol, Siena, Italy; CNR Ist Nazl Ott, Via Moruzzi 1, I-56124 Pisa, Italy; Univ Salento, Dipartimento Matemat & Fis Ennio De Giorgi, I-73100 Lecce, Italy.

Abstract: A recently introduced tuning-dressed scheme makes a Bell and Bloom magnetometer suited to detect weak variations of a radio-frequency (RF) magnetic field. We envisage the application of such innovative detection scheme as an alternative (or rather as a complement) to RF atomic magnetometers in electromagnetic-induction-imaging apparatuses. (C) 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

Journal/Review: OPTICS EXPRESS

Volume: 29 (23)      Pages from: 37081  to: 37090

KeyWords: Tomography; Squid; Probe
DOI: 10.1364/OE.437930

ImpactFactor: 3.833